Operation Manual

OmegaGTPlus / BetaGTPlus Appendix A
94
METREL GmbH VDE tester test type card
Code Autotest sequence name and descriptions Limits Barcode
01
702_Kl
_1_Iso
Testing according to VDE 702.
Class 1 appliance.
Insulation resistance and substitute
leakage current measurements are
applicable.
Earth bond: 0.30 Ω
Insulation: 1.00 MΩ
Sub leakage: 3.50mA
01
02
702_Kl
1_Iso_
BLT
Testing according to VDE 702.
Class 1 appliance with isolated
accessible cinductive parts.
Insulation resistance and substitute
leakage current measurements are
applicable.
Earth bond: 0.30 Ω
Insulation: 1.00 MΩ
Insulation - S: 2.00 MΩ
Sub leakage: 3.50 mA
Sub leakage - S: 0.50 mA
02
03
702_Kl
_1_Ia
Testing according to VDE 702.
Class 1 appliance.
Prüfung für Differenzstrom wird
eingestellt.
Earth bond: 0.30 Ω
Leakage: 3.50 mA
03
04
702_Kl
_1_Ia_
BLT
Testing according to VDE 702.
Class 1 appliance with isolated
accessible conductive parts.
Prüfungen für Differenz- und
Berührungsstrom werden eingestellt.
Earth bond: 0.30 Ω
Leakage: 3.50 mA
Touch leakage: 0.50 mA
04
05
702_Kl
_2_Iso
Testing according to VDE 702.
Class 2 appliance with isolated
accessible conductive parts.
Insulation resistance and substitute
leakage current measurements are
applicable.
Insulation - S: 2.00 MΩ
Sub leakage - S: 0.50 mA
05
06
702_Kl
_2_Ibs
Testing according to VDE 702.
Class 2 appliance.
Prüfung für Berührungsstrom wird
eingestellt.
Touch leakage: 0.50 mA
06
07
702_Kl
_1_IsoI
a
Testing according to VDE 702.
Class 1 appliance.
Prüfungen für Isolation und
Differenzstrom werden eingestellt.
Earth bond: 0.30 Ω
Insulation: 1.00 MΩ
Leakage: 3.50 mA
07
08
702Kl1
_IsoIaB
LT
Testing according to VDE 702.
Class 1 appliance with isolated
accessible conductive parts.
Prüfungen für Isolation, Differenz- und
Berührungsstrom werden eingestellt.
Earth bond: 0.30 Ω
Insulation: 1.00 MΩ
Insulation - S: 2.00 MΩ
Leakage: 3.50 mA
Touch leakage: 0.50 mA
08