Operation Manual

MI 3310 / MI 3310A SigmaGT Appendix A – Preprogrammed autotests
121
Pre-programmed autotest sequences table (cont’d)
Autotest shortcut code 09 10 11 12
Cl_2_IsoIbs Cl_2 Cl_3_Iso Cl_3
Visual test
Earth bond
Output
Limit
Time
Insulation
Output
Limit
Time
Insulation
(probe)
Output 500 V
500 V
Limit
2.00 M
0.250 M
Time 5 s
5 s
Sub leakage
Output
Limit
Time
Sub leakage
(probe)
Output
Limit
Time
Differential
leakage
Output
Limit
Time
Touch leakage
Output 230 V
Limit 0.50 mA
Time 180 s
Power
Output 230 V 230 V --- ---
Limit
Time 180 s 180 s 180 s 180 s
TRMS clamp
current
Output
Limit
Time
Polarity test