Operation Manual
MI 3310 / MI 3310A SigmaGT Instrument description
13
Warning!
These sockets are intended only for the connection to de-energized
devices.
12 Test socket
Warning!
Dangerous voltage is present on the test socket during the measurement.
Maximum output current is 16 A, test only devices with maximum rated
supply current no higher than 16 A!
Note:
For devices incorporated high reactive loading, e.g. motor with rated power >
1.5 kW, it is recommended to start measurement first and to turn on the tested
device later.
13 IEC appliance connector for testing supply cords
Warning!
The connector input is for test purpose only; do not connect it to the
mains supply!
2.2 Connector panels
Left side connector panel
14 Two T16 A / 250 V fuses for instrument protection
15 Mains supply connector
16 Battery compartment cover
17 Fastening screw for battery compartment cover
Warning!
Disconnect all accessory and tested equipment before opening the
battery cover!