Measurement of substitute leakage current of class I DUT
Measurement of substitute leakage current of fixed installed DUTs of class I
Substitute leakage measurement procedure
Select the Substitute leakage function.
Set test parameters.
Connect device under test to the instrument (see figures above).
For fixed equipment:
Disconnect mains supply of the fixed equipment;
Connect LN test socket of the instrument to L/N terminals of the fixed equipment;
Connect PE test socket of the instrument to metallic enclosure of the fixed
equipment.
Press the START key for measurement.
Example of substitute leakage current measurement results
Displayed results:
Main result ............. substitute leakage current