Operation Manual

MI 3310 / MI 3310A SigmaGT Autotest sequences
98
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
REPEAT (F3) Repeats the touch leakage current measurement.
SKIP (F4) Skips touch leakage current measurement.
6.5.9 Polarity test
Measurement is described in chapter 5.2.8 Polarity test.
If the polarity test fails or was skipped other tests will not be carried out because of
safety.
Polarity test menu
Keys:
START
Starts the polarity test.
Proceed to the next autotest sequence measurement.
REPEAT (F3) Repeats the polarity test.
SKIP (F4) Skips polarity test.
6.5.10 TRMS current measurement using clamp current adapter
Measurement is described in chapter 5.2.9 Clamp current measurement.
If the current clamp test fails or was skipped other tests will not be carried out because
of safety.
Clamp current menu
Keys:
START
Starts the TRMS current measurement.
Proceeds with the next TRMS current measurement (in continuous
measurement mode only).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).