Operation Manual

MI 3310 / MI 3310A SigmaGT Autotest sequences
104
Insulation LN - W autotest menu
Keys:
START
Starts the insulation resistance measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
REPEAT (F3) Repeats the insulation resistance measurement.
SKIP (F4) Skips insulation resistance measurement.
HELP
Displays the insulation resistance test help screen.
ESC
Ends the autotest sequence.
6.6.6 Insulation resistance (supply circuit of class II equipment to
accessible surfaces)
Measurement is described in chapter 5.3.5 Insulation resistance (supply circuit of class
II equipment to accessible surfaces). If this insulation test fails or was skipped other
tests will not be carried out because of safety.
Insulation LN - P autotest menu
Keys:
START
Starts the insulation resistance measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).