Operation Manual
MI 3310 / MI 3310A SigmaGT Appendix C – Country notes
133
Pre-programmed autotest sequences table (cont’d)
Autotest shortcut code
06 07 08 09 10
Description
KL1+2 HASPEL 5 M
2.5 MM
HASPEL 15 M
2.5 MM
HASPEL 25 M
2.5 MM
HASPEL 50 M
2.5 MM
Visual test
Earth bond
Output 200 mA 200 mA 200 mA 200 mA 200 mA
Limit
1.00 0.30 0.50 0.70 1.00
Time 5 s 5 s 5 s 5 s 5 s
Insulation
Output 500 V 500 V 500 V 500 V 500 V
Limit
1.00 M 1.00 M 1.00 M 1.00 M 1.00 M
Time 5 s 5 s 5 s 5 s 5 s
Insulation
(probe)
Output 500 V
Limit
2.00 M
Time 5 s
Sub leakage
Output
Limit
Time
Sub leakage
(probe)
Output
Limit
Time
Differential
leakage
Output 230 V
Limit 0.5 mA
Time 10 s
Touch
leakage
Output 230 V
Limit 0.25 mA
Time 10 s
Power
Output 230 V
Limit
Time 10 s
TRMS clamp
current
Output
Limit
Time
Polarity test
normal normal normal normal
RCD test
Mode
Type
Id RCD
Multi
Fase
Test