Operation Manual
MI 3310 / MI 3310A SigmaGT Measurements
55
Test circuits for substitute leakage –S probe measurement
Measurement of substitute leakage current of class II DUT
Measurement of substitute leakage of accessible isolated conductive parts of fixed
installed DUTs
Substitute leakage – S probe measurement procedure
Select the Substitute leakage-S probe function.
Set test parameters.
Connect device under test to the instrument (see figures above).
Connect EB/S probe to accessible conductive parts of the DUT.
For fixed equipment:
Disconnect mains supply of the fixed equipment;
Connect LN test socket of the instrument to L/N terminals of the fixed equipment;
Connect EB/S probe to accessible conductive parts of the fixed installed DUT
Press the START key for measurement.
Example of substitute leakage S current measurement results