Operation Manual
MI 3310 / MI 3310A SigmaGT Measurements
58
Test circuits for touch leakage current measurement
Measurement of touch leakage current
Measurement of touch leakage current on a fixed installed DUT
Touch leakage current measurement procedure
Select the Touch leakage function.
Set test parameters.
Connect EB/S probe to accessible conductive parts of the DUT (see figures above).
For Portable appliance/device:
Connect device under test to the instrument.
For fixed equipment:
Power on the fixed equipment;
Press the START key for measurement.
Examples of touch leakage current measurement results
Displayed results:
Main result ............. touch leakage current