Operation Manual
MI 3310 / MI 3310A SigmaGT Measurements
62
Test circuit for clamp current measurement
Connecting current clamp to the instrument
Clamp current measurement procedure
Select the Clamp current function.
Set test parameters.
Connect the current clamp to the instrument (see figure above).
Embrace wire(s) that has to be measured with current clamp.
Press the START key for measurement.
Examples of clamp current measurement result
Displayed results:
Main result ............. clamp current
Notes:
When measuring leakage currents, the neighboring magnetic fields and capacitive
coupling (especially from the L and N conductors) can disturb the results. It is
recommended that the clamp is as close as possible to the grounded surface and
away from wires and other objects under voltage or carrying current.
METREL offers high quality current clamps for this application.
Green socket is intended for current clamp shield terminal, if exists. This will improve
measurement of leakage current. The socket is connected to internal grounding
system and through this to PE.