Operation Manual

MI 3310 / MI 3310A SigmaGT Measurements
67
Test parameters for the Functional test
OUTPUT System voltage [230 V]
TIME Measuring time [2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s, none]
Circuit for the functional test
Functional test
Functional test
procedure
Select the Functional test function.
Set measuring time.
Connect tested DUT to the instrument (see figure above).
Press the START key for measurement.
Example of apparent power measurement result
Displayed result:
Main result ............. apparent power
Notes:
During the test, a mains voltage is connected to the DUT. If DUT contains moving
parts, make sure that it is safely mounted or protected to prevent possible danger to
the operator or damage to the DUT or surrounding environment!
Consider any displayed warning before starting measurement!