Operation Manual
MI 3310 / MI 3310A SigmaGT Autotest sequences
84
SAVE (F1) Confirms saving custom autotest sequence under entered
name.
UNDO (F2) Discards modifications and recover original entry.
ESC
Returns back to custom autotest menu.
6.2.2 Deleting an existing custom test sequence
Delete selected custom autotest sequence
Keys:
Y/ N
Confirms or rejects deleting of selected custom autotest sequence.
Any other key Return back to custom autotest menu without changes.
Note:
First custom autotest from the list cannot be deleted!
6.3 Project autotests
The Project autotests is an unique tool that dramatically simplifies and speeds up
repeated (periodic) testing of DUTs.
The main idea is to re-use known stored data (either in instrument or on a PC) of the
tested DUT. The following data can be recalled from the instrument’s memory and re-
used:
Test sequence
If the sequence has not changed (this is usually the situation) the user does not need to
care about setting the right test sequence and parameters.
DUT data
ID number, names, descriptions, addresses, comments are not needed to be re-entered
again. Old data are offered by default.
Old test results
New Project Autotest test results can be compared with previous results. The
instrument automatically calculates the trends for each measurement.
Note:
If the test results are close to the limit they should be compared with old test results.
If the trends are deteriorating, the safety of the DUT and the time between tests
should be reassessed. If the results stay stable the DUT can generally be treated as
safe.
Old test results can be uploaded from a PC back to the instrument. This brings further
advantages: