Operation Manual

MI 3310 / MI 3310A SigmaGT Autotest sequences
95
Keys
START
Starts the insulation resistance measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
REPEAT (F3) Repeats the insulation resistance measurement.
SKIP (F4) Skips insulation resistance measurement.
6.5.4 Insulation resistance – S probe measurement
Measurement is described in chapter 5.2.3 Insulation resistance – S probe.
If this insulation test fails or was skipped other tests will not be carried out because of
safety.
Insulation resistance – S probe menu
Keys
START
Starts the Insulation resistance – S probe measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
REPEAT (F3) Repeats the insulation resistance – S probe measurement.
SKIP (F4) Skips insulation resistance – S probe measurement.
6.5.5 Substitute leakage current measurement
Measurement is described in chapter 5.2.4 Substitute leakage current.
If the substitute leakage test fails or was skipped other tests will not be carried out
because of safety.