User`s manual
 MTEK6000 SERIES USER'S MANUAL 
C - 2   January 2002 
Base Temperature 
The Base Temperature parameter appears as a 
factor in the Corrected Volume calculation. It 
is one of the factors used to correct the 
flowing volume, as registered by the meter 
itself, to the base volume used for calculating 
“Standard Volume”. The default value of this 
parameter is 60.000 degrees F. 
Calibrate Mode Time-out 
The fractional portion of this parameter is the 
Calibrate Mode Time-out. It controls how 
long the unit will remain powered up in the 
calibration mode if no key is being pressed.  
For example, a calibration time-out of 30 
minutes would be entered as XX.30. The 
default value is 60.30. 
Calibration Password 
A password can be entered to prevent 
unauthorized users from accessing calibration 
mode. The value may be up to six digits long 
(including the decimal point and sign). The 
Default value is 0. 
Configuration Password 
A password can be entered to prevent 
unauthorized users from accessing 
configuration mode. The value may be up to 
six digits long (including the decimal point 
and sign). The Default value is 0. 
Corrected Pulse Output -Cubic Unit / Pulse 
This parameter should reflect the value of the 
corrected pulse output sent from the station to 
an external device. For example, if each 
output pulse represents 1000 cubic feet, then 
this parameter should be set to 1000. The 
Default value is 1000. 
Corrected Pulse Out Enable/Disable 
Corrected pulses can be sent to an external 
device by enabling the Pulse Output Channel.  
The rate at which pulses are generated is 
determined by the Corrected Pulse Out - CF 
per Pulse parameter. To enable Relay #1 for 
pulse output, enter a value of 35.8. Enter -1 
to disable. The Default value is -1. 
Corrected Volume 
The corrected volume is calculated based 
upon AGA report #7 or #8 and reflects the 
corrected volumetric flow taking the base 
conditions into consideration. The Initial 
value is 0. 
Corrected Volume (Pressure Only) 
The corrected volume is calculated based 
upon AGA report #7 and reflects the 
corrected volumetric flow taking only the 
Pressure base conditions into consideration. 
The temperature factors have no effect. The 
Initial value is 0. 
Corrected Volume and Flow Multiplier 
This multiplier specifies the output value for 
corrected volume and flow. For example, if 
the device is to measure flow in thousands of 
cubic feet (MCF), select Thousands (1000's) 
of units from the pick list. The Default value 
is 1000. 
Ten Thousands (10,000's) of units  10000 
Thousands (1000's) of units   1000 
Hundreds (100's) of units   100 
Tens (10's) of units     10 
Single (1's) units     1 
Counter Fault Monitoring (EFC Only) 
This parameter can be used to enable or 
disable fault monitoring on the main counter 
#1. Form C (3-wire) input is required for this 
function. In this mode, two working counters 
are monitored for discrepancies. A “Delta” 
value will be accumulated which reflects any 
discrepancies between the counters. If any of 
the dual-reed switches should be defective, 
and the “Delta” exceeds the Counter Fault 
Threshold parameter, the input pulses will 
automatically switch to either working single- 
ended counter. The EFC will generate a 
Faulty Counter alarm. To enable, set this 
parameter to 1 and to disable set to 0. The 
default value is 0 for disable. 