MX 26 MULTIMETRE NUMERIQUE DIGITAL MULTIMETER DIGITAL MULTIMETER MULTIMETRO DIGITALE MULTIMETRO DIGITAL User's manual Copyright © page 15 Chapter II 906129650 - Ed.
Interface optique RS232 RS232 optical interface optische Schnittstelle RS232 Interfacia ottica RS232 Interfaz optica RS232 5 19 6 20 18 MX 26 21 17 16 15 22 14 13 8 12 11 9 7 10 COM 1 2 3 4 Multimètre digital portable 5000 points, avec entrées courant
LEGENDE / LEGEND / BESCHREIBUNG / LEGENDA / LEYENDA 1 2 3 4 5 6 7 8 9 10 11 Borne d’entrée (positions 12 à 20) Entrée de référence du multimètre Borne d’entrée: gammes 500 mADC et 500 mAAC+DC Borne d’entrée: gammes 10 ADC et 10 AAC+DC Changement de gamme: RANGE Fonctions secondaires: AC+DC Mesure de pics +/Gel de l’affichage Activation du rétro-éclairage Mémorisation automatique Mise hors tension 12 13 14 15 16 17 18 19 20 21 22 Mesure de tensions AC: gammes 5 VAC à 750 VAC (basse impédance) Mesure de te
Chapter II USER’S MANUAL CONTENTS 1. GENERAL INSTRUCTIONS ...................................................................................................................16 1.1. Precautions and safety measures ...................................................................................................16 1.1.1. Preliminary...................................................................................................................................16 1.1.2. During use .........................
Chapter II 1. GENERAL INSTRUCTIONS You are the new owner of a 5000 ct portable digital multimeter and we thank you for your choice. This instrument complies with the specification set out in the IEC 61010-1 + A1 + A2, concerning safety requirements for electronic measuring apparatus. To get the best service from this instrument, read carefully this user's manual and respect the detailed safety precautions. The contents of this manual must not be reproduced in any form whatsoever without our consent. 1.1.
Chapter II ∗ When the multimeter is linked to measurement circuits, do not touch unused terminals. ∗ When the range of the value to be measured is unknown, check that the range initially set on the multimeter is the highest possible or, wherever possible, choose the autoranging mode. ∗ Before changing functions, disconnect the test leads from the circuit under test.
Chapter II 1.2. Protection mechanisms This instrument is fitted with various protection mechanisms : ∗ Varistor protection for limiting transients of over 1100 V at the VΩ terminal, particularly 6 kV pulse streams as defined by the French standard IEEE 587. ∗ A PTC (positive temperature coefficient) resistor protects against permanent overvoltages of up to 600 V during resistance, capacitance, continuity and diode test measurements. This protection is reset automatically after overload.
Chapter II 2. DESCRIPTION This compact multimeter is a self-contained with an appropriate mechanical construction, enables hand-held version, with a protective elastomer case. It is designed for a high degree of user safety, maximum protection and unrivalled performance. 2.1. Selector switch This multimeter is a standalone, hand-held professional measuring instrument, capable of measuring the following quantities (accessed by the twelve-position rotary selector switch): ∗ ∗ ∗ ∗ ∗ ∗ ∗ ∗ ∗ ∗ ∗ 2.2.
Chapter II 3. GETTING STARTED 3.1. Connecting the test leads Connect the black lead to the COM socket (for all measurements). Depending on the position of the selector switch, connect the red lead as follows: Rotary selector switch position VLOW Z, VAC+DC, VDC, ADPDC (ADPAC+DC), Hz, 500 mADC (500 mAAC+DC) 10 ADC (10 AAC+DC) Input terminal , Ω, , VΩ 500 mA 10 A RS232 lead : Connect the RS232 optical interface (optional kit) to the top of the multimeter casing and the DB9F plug to PC (COM port).
Chapter II 3.5.2. Battery self-test When the BAT indication appears on the display, the instrument still has approximately 12 hours of operation, but specifications can no longer be guaranteed. Replace the battery. 3.5.3. Replacing the battery or fuses * Caution ! Disconnect test leads from any circuit under test, turn the meter off and remove test leads from the input terminals . Use the following procedure: 1 - Disconnect test leads from any input terminals.
Chapter II 4. FUNCTION DESCRIPTION 4.1. RANGE / AC+DC key The RANGE key is operative in following positions of the selector switch : VLOW Z, VAC+DC, VDC, Ω, , Hz. This key makes it possible : • In AUTO (Autoranging) mode to switch to MANUAL mode (short press). • In MANUAL mode, to select the next range (short press) or return to AUTO mode (long press), the AUTO sign appears on the display. This can be used to switch on the multimeter again after an automatic shutdown.
Chapter II 4.1.3. 10 ADC / 10 AAC+DC Position 10 ADC DC current measurement (Range 10 ADC) AC+DC key 10 AAC+DC AC+DC current measurement (TRMS) (Range 10 AAC+DC) The AC+DC sign appears on the display. AC+DC key 4.2. +/- Peak key The fast positive or negative peak measurement functions (> 1 msec.) can be accessed by repeatedly pressing this key in the VDC, ADPDC, mADC and 10 ADC functions. This key also desactivates the auto power-off, if it is pressed when switching on (“P-OFF” is displayed).
Chapter II 5. SOFTWARE KIT (optional) Connection of the kit and installation of the SX-DMM software : With the software kit option, the multimeter can interface directly with a computer or PC: 1. Connect the optical plug at the top of the shockproof sheath. A mechanical fail-safe device is provided on the left-hand side to prevent connection the wrong way round. 2. Plug the connector into one of the PC's "COM" ports. 3.
Chapter II 6. TECHNICAL SPECIFICATIONS 6.1. General Only the values assigned tolerances or given limits are guaranteed values. Values without tolerances are given for information only, without guarantee (standard NFC 42670). The measurement errors must be considered in an environment of reference temperature (refer to § 6.2.11). It is essential that all measuring instruments are regularly calibrated. 6.2. Characteristics Accuracy is ± [% reading (R) + number of digits (D)].
Chapter II 6.2.3. DC current Selector switch position Ranges Accuracy Max volt. drop Protection Fuses (∗) Resolution 500 mADC 500 mADC 0.3 % R + 2 D < 600 mV 600 VRMS F1 + F2 100 µADC 1%R+2D < 700 mV 600 VRMS F2 10 mADC 10 ADC 10 ADC (∗∗) (∗) Refer to fuse specifications section 3.5.3. (∗∗) Acceptance 20 A overload during 30 s. max, observing a 5 min break at least between 2 measurements. Beyond 7 A, the measurement is limited to one minute.
Chapter II 6.2.6. Capacitance ) Note Discharge all capacitors before taking measurements. Selector switch position Ranges Accuracy Measurement current 50 nF 100 nA 500 nF 1 µA 5 µF 10 µA 50 µF 1 % R + 2 D 100 µA 500 µF 1 mA 5000 µF 1 mA 50 mF 2 % R + 2 D 1 mA Max measurement Protection Resolution time (∗) <1s 10 pF <1s 100 pF <1s 1 nF 10 nF <1s 600 VRMS <2s 100 nF 1 µF ≈ 3 s/mF 10 µF ≈ 3 s/mF (∗) Overload protection can be reset automatically.
Chapter II 6.2.10. General information Mechanical Dimensions Weight (inc. battery) Casing and circuit 170 x 80 x 35 mm 285 g self-extinguishing materials Packaging Dimensions Weight 230 x 155 x 65 mm 385 g Power supply Power requirements Low battery indication single 9 V alkaline battery (6LF22) BAT is displayed when the battery voltage drops below operating voltage 500 hours approx.
Chapter II 6.3. Accessories 6.3.1. Supplied with the multimeter One set of test leads with safety probes One 6LF22 9 V battery One user’s manual Protective elastomer case with RS232 window AL0042 906129650 HX0010 6.3.2. Optional Communication kit Acquisition Software (part. nr. SX-DMMC) RS232 lead (DB9F) (part. nr.