Operation Manual
Chapter II
5000-count portable digital multimeter, with current inputs
25
6. TECHNICAL SPECIFICATIONS
6.1. Genera
l
Only the values assigned tolerances or given limits are guaranteed values. Values without
tolerances are giv
en for information only, without guarantee (s
tandard NFC 42670). The
measurement errors must be considered in an env
ironment of reference temperature (refer to
§
6.2.11).
It is essential that all measuring instruments are regularly calibrated.
6.2. Characteristics
Accuracy is ± [% reading (R) + number of digits (D)].
{Accuracy : ‘’n %R + n D’’ means ‘’n % of the reading + n digits’’ }
6.2.1. DC voltages
Selector switch
position
Ranges
Accuracy
Input impedance
Protection
(∗)
Resolution
ADPDC 500 mVDC
10 MΩ
0.1 mVDC
5 VDC
11 MΩ ± 1100 V
PEAK
1 mV
DC
VDC 50 VDC 0,3 % R + 2 D 775 Vrms 10 mVDC
500 VDC
10 MΩ
100 mVDC
1000 VDC 1 VDC
(∗) Acceptance permanent max. voltage
Number of points : 5 000
Range selection : automatic or manual for the 5 V, 50 V,
500 V, 1000 V ranges
Common mode rejection : at 50 and 60 Hz, better than 120 dB
Serial mode rejection : at 50 and 60 Hz, better than 60 dB
Intermittent signal sounds and « OL » on the display in case of range overpassing.
For measurements performed on alternative signals, the selected range must tally with the max.
value of the signal peak.
Additional error in Pk +/- mode for a pulse width of ≥ 1 ms : 1 % R ± 50 D
6.2.2.
AC voltages (AC and AC+DC)
Switch Range Input Protection Accuracy Resolution
position impedance (*) 40Hz to 1 kHz 1 to 4 kHz 4 to 20 kHz 20 to 50 kHz 50 to 100 kHz
5 VAC 1 mVAC
VLOW Z 50 VAC
500 kΩ
± 1100 Vpk 1 % L + 3 D 1 % L + 3 D 2 % L + 3 D 3 % L + 3 D 5 % L + 3 D 10 mVAC
500 VAC 775 Vrms 100 mVAC
750 VAC - - - - 1 mVAC
ADPAC+DC 500 mVAC+DC
10 MΩ
//100 pF
± 1100 Vpk
775 Vrms
1.5 % L + 3 D - - - - 0.1 mV
AC+DC
5 VAC+DC
11 MΩ
1 mVAC+DC
VAC+DC 50 VAC+DC ± 1100 Vpk 1 % L + 3 D 1 % L + 3 D 2 % L + 3 D 3 % L + 3 D 5 % L + 3 D 10 mVAC+DC
500 VAC+DC
10 MΩ
775 Vrms 100 mVAC+DC
750 VAC+DC - - - - 1 VAC+DC
(∗) Acceptance permanent max. voltage
Specifications applicable from 10 % to 100 % of the range for a sinusoidal signal in band 40 Hz
to 20 kHz and from 20 % to 100 % of the range in band 20 kHz to 100 kHz.
Number of points : 5 000
Range selection : automatic or manual for the 5 V, 50 V,
500 V, 750 V ranges
Common mode rejection : at 50 and 60 Hz, better than 60 dB
Additional error according to crest factor : 0.5 % for a crest factor of 2 to 3
1 % for a crest factor of 3 to 6
Intermittent signal sounds and « OL » on the display in case of range overpassing.