3.4 Multipoint calibration
IC Metrodata for Win95
31
Enter concentrations for calibration
• Select Open method of the File menu.
• Select the previously saved method in the
File Open window and
click
OK.
• Select
Calibration / Concentrations of the Method menu.
• Click the
Add button and confirm the suggested number 2 for the
Level in the Add level window by clicking OK.
• Enter the concentrations of all components of the second standard
in the new
Level 2 column.
• Repeat this procedure for every further standard.
• If all standards have been entered, close the Concentrations win-
dow by clicking
OK.
• Close the empty chromatogram window. The message
Method
*.mtw was modified. Save changes?
appears.
• Click the
Yes button. The File Save As window appears.
• Click the
OK button. A window appears with the message c:\..... al-
ready exists. Overwrite?
• Click the OK button.