3.4 Multipoint calibration
IC Metrodata for Win95
35
Save method
Select Save method as of the File menu. The File Save As window
appears.
Click the OK button. A window appears with the message c:\..... al-
ready exists. Overwrite?
Click the OK button.
Record a sample chromatogram
Click on
or select Restart method of the Acquire menu.
Enter a sample name (max. 14 characters, e.g. "drinking water") as
Ident in the Start Run window.
If required, enter more parameters (e.g.
Dilution) to characterize
sample and acquirement conditions.
Click the
OK button.
The window for the on-line display of the chromatogram opens.
The status display below shows the message
Waiting.
Set the injection valve at the 733 IC Separation Center to "FILL"
(manually or by program) and fill the sample loop with sample.
Set the injection valve at the 733 IC Separation Center to "INJECT"
(manually or by program). The data acquisition is started automati-
cally and stopped as the analysis time has passed.