Owner manual

4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-24
2
Storage on the data card
A data card is included in the standard equipment of the 746 VA Trace
Analyzer for the storage of methods and determinations. Before first-time
use in the 746 VA Trace Analyzer, this data card must first be prepared.
The steps required for this (insertion of the battery, affixing the label, for-
matting, plugging in) are described in section 3.2.6.
Insert the prepared and formatted data card in the 746 VA Trace Ana-
lyzer and press the yellow <DATA CARD> key. The dialog page "DATA
CARD" appears showing the methods and determinations stored on the
card:
Files : 0 of 256 DATA CARD Capacity: 131072 Bytes
Available: 1353520 Bytes Label :
File
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
Modify
column
Modify
line
Delete
file(s)
Copy
to
Copy
from
Unselect
file(s)
Change
display
>>
The header of the dialog page "DATA CARD" contains information on the number of
files stored and the available storage space. This is followed by the names of all stored
files displayed in 4 columns.
You will find further information regarding this dialog page in section 5.4.3.
Press the softkey <Copy from>. The softkey bar changes and presents
the sources for copying a method or determination to the data card.
Press the softkey <Working storage> (determination in the working
storage). The determination loaded in the working storage is copied to
the data card, the file name (in this example 03201614.dtm) appears in
the first column under the title File.
Press the softkey <Change display>. The type of display changes, in
addition to the name of the determination a field for content information
and the storage space needed for the determination in bytes appear.
Both file name and content information can be changed for improved
documentation. To do this, press the softkey <Modify line> to enter the
edit mode. The first 8 characters of the determination name can now be
changed in any way (the extension .dtm must be retained). Now press
the <ENTER> key and enter any comment in the Content field for the
determination (e.g. Lead determination with method TESTPb of
20.03.1997). After confirmation with the <ENTER> key, you are again in