Owner manual

5.6.3 SUBSTANCES
746 VA Trace Analyzer / 747 VA Stand
5-135
Parameter Range; Default value / Meaning
Display / Plot display only
Display/plot (for measurement mode PSAMODE)
Title for those parameters which concern the display and plotting of
substance curves recorded with the measurement mode PSAMODE.
t/U.scale auto, fixed; auto
Scaling of the y axis of PSA curves
The following two possibilities are available for scaling of the y axis of
PSA curves for the display and plotting of substance curves:
auto Automatic scaling with matching to the measured
experimental values
fixed Fixed scaling with the preset values t/U.begin and
t/U.end
If all measurement values are outside the y axis fixed with t/U.scale,
the y axis for printing the substance curve is scaled automatically. In
this case, the message scale auto is printed in the header line.
t/U.begin -999.99...999.99 ns/V...s/V;
Initial value for scaling of the y axis of PSA curves
This parameter appears only if t/U.scale = fixed has been se-
lected.
If no value is entered for t/U.begin, the initial value is determined
automatically.
t/U.end -999.99...999.99 ns/V...s/V;
Final value for scaling of the y axis of PSA curves
This parameter appears only if t/U.scale = fixed has been se-
lected.
If no value is entered for t/U.end, the final value is determined auto-
matically.
U.div 10 mV/cm...50 V/cm; 50 mV/cm
Division of the voltage axis
Division of the voltage axis in mV/cm for the plotting of substance
curves (not effective in display of the curves).
U.begin -3000...3000 mV;
Initial value for scaling of voltage axis
If no value is entered for U.begin, the initial voltage U.start of the
sweep is taken as the initial value.
U.end -3000...3000 mV;
Final value for scaling of voltage axis
If no value is entered for U.end, the final voltage U.end of the sweep
is taken as the final value.
t/U.end
t/U.begin
U.begin U.end