5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-142
Method: XXXXXXXX SUBSTANCES
........ - XXXXXXXX
Calibration XXXX-XX-XX XX:XX Coefficients
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
Technique std.add. Y.reg XXXXXXXXXX
Curve type ......... Slope XXXXXXXXXX
Nonlin. XXXXXXXXXX
Mean dev. XXXXXXXXXX
Additions
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
Soln.name ........ ........ ........ ........
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ
Mass conc. .........g/L .........g/L .........g/L .........g/L
Range min .........g/L .........g/L .........g/L .........g/L
Range max .........g/L .........g/L .........g/L .........g/L
M.conc./cm .........g/L .........g/L .........g/L .........g/L
Modify
column
Modify
line
Select
subst.
Link to
segment
Delete
subst.
Subst.
name
>>
Softkeys Meaning
Meaning of the softkeys F1...F7, see page 130
>>
Switching to subpage "Substance parameters"
Switching to the subpage "Substance parameters" with the parame-
ters for substance recognition, baseline calculation, scaling and
evaluation.
Parameter Range; Default value / Meaning
Calibration
XXXX-XX-XX
XX:XX
display only
Date of calibration
Date and time of the calculation of the standard addition.
Technique none, std.add., cal.crv., int.std.; none
Calibration technique = Standard addition (std.add.)
With the standard addition method, a known quantity of the substance
under investigation is added to the sample one or more times (see
section 6.9.4). The addition can be performed manually with a pipette
or the 6.5611.010 pipetting equipment, automatically with a 685
Dosimat or 700 Dosino attached to the 746 VA Trace Analyzer or with
the 695 Autosampler. There are two types of instructions for this
addition: