Owner manual

5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-152
Parameter Range; Default value / Meaning
Mean dev. -999.9...999.9 e-20...e+20;
Mean deviation of calibration curve
Display of the calculated, mean deviation of the measured values
about the calibration curve in A, W or s. This value can be changed
manually.
The size of Mean dev. can be used as a measure of the quality of a
calibration curve.
This value can be used in formulae as KD:<subst> (<subst> = sub-
stance name) for further calculations.
CSoln.name 8 characters;
Name of calibration solutions
The entry of a name for the calibration solutions is optional. The
names are used as follows:
Without 695 Autosampler, before every manual addition of a
calibration solution to the measuring vessel the message Next
Standard, CSOLN 'CSOLN.name''Mass conc.' > M appears with
details of the name and the mass concentration of the calibration
solution.
If work is performed with the 695 Autosampler with Auto.samples
= on, all calibration solutions must be entered on the "SAMPLES"
page. We advise you to enter the names CSoln.name there as
sample identification.
Mass conc. 0.001...999.99 ng...kg/L;
Mass concentration ρρ
n
of calibration solutions
The mass concentration must be specified for all n calibration solu-
tions defined by the instruction CAL)n.