Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-189
The phase dependent differences in the current measurements give
curves which can be evaluated using the baselines of type AC2 rel. or
AC2 abs. (see section 5.6.3).
U / 2
I.abs
I
U
Comments The AC2MODE measurement mode can be used with all types of
electrodes together with the instructions MEAS, SWEEP, DSWEEP,
FSWEEP or TSWEEP.
The electrode type DME, SMDE, HMDE or RDE must be entered before
the AC2MODE measurement mode.
With the AC2MODE measurement mode, the potentiostat always
operates in the operating mode "fast". SWEEP and FSWEEP are thus
identical for this mode.
The quantity t.drop important for the definition of t.step depends
on the type of electrode selected:
DME t.drop = 20 ms
HMDE and RDE t.drop = 0 ms
SMDE t.drop = Drop.size ×× 40 ms
see also AC1MODE, SWEEP, DME, SMDE, HMDE, RDE
Σ(I
n+
– I
n–
)
n
I =
I.rel = I
f
+ I
r
I
f
I
r
n = Meas.cycles