Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-196
A>M Page: O, S Stand: A Abbreviation: A>
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. A>M ..... addition AUX -> MEAS V.add ........mL
Function Transfer solution from auxiliary vessel (AUX) on the 695 Autosampler to
the measuring vessel (MEAS) on the 747 VA Stand.
Parameter Range; Default value / Meaning
V.add 0.010...50.000 mL; 0.100 mL
Transfer volume
Solution volume that should be transferred from the auxiliary vessel on
the sample rack of the 695 Autosampler to the measuring vessel on the
747 VA Stand.
Description The A>M instruction is used to transfer the solution volume specified
under V.add from the auxiliary vessel (AUX) on the 695 Autosampler to
the measuring vessel (MEAS) on the 747 VA Stand.
Comments The A>M instruction can be used only in operation with the 695
Autosampler.
If the solution volume in the auxiliary vessel < V.add, an error
message appears.
see also F>A, F>M