5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-219
(CSWEEP Page: S Stand: M, A Abbreviation: (CS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. (CSWEEP
Function Start of the block for a cyclic voltage sweep.
Description The (CSWEEP instruction opens a block in a segment in which the ramps
for the cyclic voltage sweep are defined. This block can contain
maximum 5 RAMP instructions, other instructions are not allowed. The
block must be closed with the CSWEEP) instruction.
Comments • Only one CSWEEP block is allowed per segment.
• CSWEEP blocks may not be located within REP loops.
• The electrode type (admissible are HMDE or RDE) and the measure-
ment mode CYCMODE must be entered before the CSWEEP block.
• The number of measurement cycles for the CSWEEP block (max. 9)
is defined in the CYCMODE instruction by the Meas.cycles parameter.
• Within a CSWEEP block a wait time between two ramps can be
programmed only by insertion of an additional, slow ramp as flat as
possible.
Example see RAMP
see also CSWEEP), CYCMODE, RAMP