5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-219
(CSWEEP Page: S Stand: M, A Abbreviation: (CS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. (CSWEEP
Function Start of the block for a cyclic voltage sweep.
Description The (CSWEEP instruction opens a block in a segment in which the ramps
for the cyclic voltage sweep are defined. This block can contain
maximum 5 RAMP instructions, other instructions are not allowed. The
block must be closed with the CSWEEP) instruction.
Comments Only one CSWEEP block is allowed per segment.
CSWEEP blocks may not be located within REP loops.
The electrode type (admissible are HMDE or RDE) and the measure-
ment mode CYCMODE must be entered before the CSWEEP block.
The number of measurement cycles for the CSWEEP block (max. 9)
is defined in the CYCMODE instruction by the Meas.cycles parameter.
Within a CSWEEP block a wait time between two ramps can be
programmed only by insertion of an additional, slow ramp as flat as
possible.
Example see RAMP
see also CSWEEP), CYCMODE, RAMP