5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-251
Description The PADD>M instruction is used for the automatic or manual addition of
standard addition solutions. In contrast to the ADD>M instruction, the
standard addition volume V.add can not be preallocated, but is
recalculated in every measurement from the evaluation value EV:<subst>
measured beforehand for any substance and a proportionality coefficient
Prop.coeff..
The automatic addition can be performed with a 685 Dosimat, a 700
Dosino or directly with the 695 Autosampler. The appropriate position of
the standard addition solution must be specified on the "SOLUTIONS"
page (see Soln.name)
Comments The PADD>M instruction must be located within an ADD loop.
In the last execution of the ADD loop, the PADD>M instruction is no
longer executed.
On execution of the PADD>M instruction with a 685 Dosimat or 700
Dosino, a wait is made until the addition is complete before the
instruction following PADD>M is executed.
Example see (CASE
see also (ADD, ADD)n, ADD/M, ADD>M