5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-285
SIMPULSE Page: O, S Stand: M, A Abbreviation: SI
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SIMPULSE .....
Function Simulation of standard additions with the 3.496.8380 VA Simulator.
Description The SIMPULSE instruction is used to put out a total of 100 pulses at the
3.496.8380 VA Simulator which cause an increase in the current
measurement and hence a simulation of standard additions. The VA
Simulator can be reset to the original current level with the SIMRESET
instruction.
Comments The 3.496.8380 VA Simulator and the associated 3.496.5190
Connecting cable are available as an option. The VA Simulator gen-
erates 2 polarographic current waves with half-wave potentials of ca.
–540 mV and ca. –1070 mV in a defined voltage range. You will find
further details in the 3.496.8380-SB Circuit Description supplied with
the VA Simulator.
To obtain a larger current increase, the SIMPULSE instruction can also
be entered twice in succession.
Example Test program for 3 simulated standard additions with the 3.496.8380 VA
Simulator:
Method SIMI .mth OPERATION SEQUENCE
Title Test program
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 SMPL/M V.fraction mL V.total L
2 SIMRESET 1.0
3 (ADD
4 SEGMENT Segm.name Seg1
5 SIMPULSE
6 ADD>M Soln.name V.add 0.100 mL
7 ADD)3
8 END
Method SIMI SEGMENT
Seg1
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 HMDE Drop size 4 Meas.cell normal
2 DPMODE U.ampl 50 mV t.meas 20.0 ms
t.step 0.20 s t.pulse 40.0 ms
3 SWEEP 50.4 U.start -100 mV U.step 6 mV
U.end -1600 mV Sweep rate 30 mV/s
4 END
see also SIMRESET