5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-296
Description The SSWEEP instruction starts the voltage sweep defined in the parame-
ters.
U
t
U.start
U.end
U.step
t.step
Comments The electrode type must be defined before the SSWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
The measurement mode must be defined before the SSWEEP
instruction. Admissible is only DCTMODE.
Apart from the SSWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
With the SSWEEP instruction maximum 1000 measured values can be
recorded.
see also SWEEP, DSWEEP, FSWEEP, DME, SMDE, HMDE, RDE, DCTMODE