Owner manual
5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-296
Description The SSWEEP instruction starts the voltage sweep defined in the parame-
ters.
U
t
U.start
U.end
U.step
t.step
Comments • The electrode type must be defined before the SSWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
• The measurement mode must be defined before the SSWEEP
instruction. Admissible is only DCTMODE.
• Apart from the SSWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
• If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
• With the SSWEEP instruction maximum 1000 measured values can be
recorded.
see also SWEEP, DSWEEP, FSWEEP, DME, SMDE, HMDE, RDE, DCTMODE