Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-297
STIR Page: O, S Stand: M, A Abbreviation: ST
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. STIR ..... Rot.speed ......../min
Function Switch on stirrer.
Parameter Range; Default value / Meaning
Rot.speed 0, 200, 400, ..., 3000 /min; 2000 /min
Rotational speed
Setting of the number of revolutions of the stirrer per min.
Description The STIR instruction is used to switch on the stirrer of the 747 VA Stand.
The stirrer comprises the 6.1246.010 Drive shaft and the 6.1204.090
Stirrer tip (see section 3.8).
The stirrer remains active until it is switched off by the ØSTIR instruction
or END in the main program on the "OP.SEQUENCE" page.
Comment If the stirrer should be switched off automatically after a definite time, the
TSTIR instruction can also be used.
see also ØSTIR, TSTIR