Owner manual

3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-18
3.4.4 Filling the capillary without vacuum
The glass capillary 109109 can normally be filled with mercury by the method de-
scribed here without vacuum. However, if difficulties regarding stability or repro-
ducibility arise with a capillary filled in this manner, try to fill the capillaries by the al-
ternative method with vacuum (section 3.4.5).
To fill the mounted glass capillary 109109 (section 3.4.3) with Hg without vacuum, pro-
ceed as follows:
1
Install multi-mode electrode in 747 VA Stand
With the measuring head arm 4040 tilted back, slide the empty measuring
vessel 2828 into the holder 2626 of the 747 VA Stand and then lower the
measuring head arm 4040.
Carefully insert multi-mode electrode 3838 in opening 7676 of the measuring
head 4141 (during insertion, the tip of the capillary 109109 must not touch the
measuring head) and push in carefully as far as it will go. Avoid water
drops touch the tip of the capillary.
2
Connect multi-mode electrode
Screw FEP tubing 5151 for the inert gas supply into connection 9393 of the
multi-mode electrode 3838.
Screw FEP tubing 5959 for the inert gas supply into connection 9494 of the
multi-mode electrode 3838.
Attach electrode cable 3737 (WE) to screw connection 110110 of the multi-
mode electrode 3838: push cable lug under the screw and then tighten
screw firmly.
3
Fill capillary with mercury
Switch on 746 VA Trace Analyzer with mains switch 1111 (746 VA Trace
Analyzer and 747 VA Stand must first be installed properly as described
in section 3.2 and section 3.3).
Select dialog page "STAND" and then switch on the inert gas supply to
the 747 VA Stand by pressing the softkey <Pressure>. This pressurizes
the multi-mode electrode 3838 and the mercury begins to flow slowly out of
the capillary.
Gently tap the MME with your finger (to remove any air bubbles) and
allow the mercury to flow out of the capillary into the empty measuring
vessel for approx. 2 min.
Fill measuring vessel 2828 with 10 mL ultrapure water and add 1 drop KCl
solution (in pure water, mercury drops from the capillary only with diffi-
culty).
Allow mercury to flow out of the capillary for ca. 2 min while checking the
drop formation: The drop time should be ca. 3 s.
4
Adjusting the sealing needle 9797
Turn slotted screw 9696 using a suitable coin slowly in a clockwise direction
until the mercury flow stops.
Open slotted screw 9696 slightly in an anticlockwise direction until the
mercury flow restarts.