5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-309
Example Variation of the pulse amplitude with the SQWMODE measurement mode:
Method: VAR SEGMENT
Seg1
================================================================================
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
1 (VAR
2 HMDE Drop size 4 Meas.cell normal
3 VARNI Param.name U.ampl Delta 0.005
4 SQWMODE U.ampl 10 mV Modul.freq. 50 Hz
t.step 0.30 s Prep.cycles 0
t.meas 2.0 ms Meas.cycles 4
5 SWEEP 21.0 U.start -700 mV U.step 6 mV
U.end -300 mV Sweep rate 20 mV/s
6 VAR)8
7 END .
In this segment, the pulse amplitude in increased by 5 mV in every
measurement from 10 mV to 50 mV. A total of 9 measurements are
performed and are designated in the results as VR = 00...80.
see also (VAR, VARNI, (REP, REP)n