Owner manual

7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-19
Error message Meaning/Causes Rectification Exit
MEASURE - ERROR 30:
Missing voltage value
Voltage value in sweep is not
defined (U.end, U.start or
U.meas)
Enter voltage values in the SWEEP
instruction
E/S
MEASURE - ERROR 31:
Illegal replication
Inadmissible replication in
conjunction with measurement
mode CYCMODE
Clear replication (REP...REP)n E/C
MEASURE - ERROR 32:
Missing mass concentra-
tion (SOLUTIONS)
The mass concentration Mass
conc. is missing on the
”SOLUTIONS” page
Enter mass concentration Mass conc.
on the ”SOLUTIONS” page
E/S
MEASURE - ERROR 33:
Standard addition with
different solutions
In the standard addition,
spiking was not performed with
the same standard addition
solution (different sequence in
conjunction with CASE instruc-
tions)
Check programming of the CASE
instructions
E/C
MEASURE - ERROR 34:
<Start> from this page
impossible
Program start from the
”STAND” and ”CURVES”
pages not possible
Switch to the ”MONITORING” page and
press <START>
E/C
MEASURE - ERROR 35:
Variation parameter out
of range
The range limits of the varied
parameter were exceeded
Change Delta parameter in the
VARNI instruction or number of variation
cycles
E/S
MEASURE - ERROR 36:
Missing solution
An entry is not available for
every substance on the
”SOLUTIONS” page
Missing or wrong substance
name
Enter solution for every substance on
the ”SOLUTIONS” page (also with
mixed solutions)
Check substance name on the
”SOLUTIONS,” SUBSTANCES” and
”OP.SEQUENCE” or ”SEGMENTS”
pages
E/S
MEASURE - ERROR 37:
Measurement data
inconsistent
The experimental data are
incomplete, e.g. owing to
instrument being switched off
during an ongoing measure-
ment
Restart measurement E/C
MEASURE - ERROR 38:
Too small volume in
FEED or AUX
Insufficient solution for the
desired transfer instruction in
the sample vessel FEED or in
the auxiliary vessel AUX on the
695 Autosampler
Lower volume V.add for transfer
instruction
E/S
MEASURE - ERROR 39:
Incomplete calibration
The calibration has not been
run to completion
Restart calibration E/S
MEASURE - ERROR 40:
PREP demands Auto.batch
= on
PREP loops are admissible only
when Auto.batch = on is set
Set Auto.batch = on on the
”MONITORING” page
E/C
MEASURE - ERROR 41:
Different V.total in
SMPL
On entry of several SMPL
instructions, V.total must be
identical in all cases
Set V.total the same for all SMPL
instructions on the ”OP.SEQUENCE” or
”SEGMENTS” page
E/C
MONITORING - ERROR 1:
Can not start as busy
measuring
The called-up function can not
be executed as a measurement
is running
Wait until measurement ended E/C
MONITORING - ERROR 2:
Live display only when
BUSY
Display of the live curve not
possible as no measurement is
running
Start measurement E
OP.SEQ. - ERROR 1:
Missing END instruction
Missing END instruction Append END instruction to end of
program on the ”OP.SEQUENCE” page
Esc
OP.SEQ. - ERROR 2:
Nesting violation: line
##
Error in the nesting of instruc-
tions with brackets
Check program on the ”OP.SE-
QUENCE” page for correct use of
instructions with brackets and correct if
necessary
Esc