Owner manual

7.4 Troubleshooting
746 VA Trace Analyzer / 747 VA Stand
7-39
Peak is in the highest µA range
The sample volume is too large: reduce the volume. Carry out the analysis
again.
The enrichment period (for HMDE) is too long: Reduce the time.
If necessary use a SMDE or DME electrode instead of HMDE.
Double peak
Check MME. If necessary, change capillary (see section 3.4.8) or replace
sealing needle (see section 3.4.9).
Organic components interfere with the analysis? Carry out a UV digestion or
other suitable sample preparation.
If a second element is present at the same potential: add this element to the
sample and carry out the analysis again. If the second peak has become
higher then the second element is present.
Might it be possible to selectively mask this second element with a complex
former?
Electrolyte solution too old: make up a new one. Its working life with organic
additives may be as short as 1 day or less.
For Cu: work without chlorides in the electrolyte or increase the chloride con-
centration massively.
Try another measurement mode like AC1. If one substance is reversible and
the second one irreversible, only the reversible substance is detected by
AC1.
Standard addition peaks displaced
If HMDE is used potential displacements of more than 2030 mV are often
normal and have to be accepted; particularly in adsorption voltammetry.
Standard solutions have been made too acidic.
Buffering capacity of the electrolyte is not sufficient: increase electrolyte vol-
ume.
No addition
Has the correct standard solution been used or is the concentration of the
solution too low: increase the volume of the standard addition or use a higher
concentration or reduce the sample amount accordingly.
If organic components are present: carry out a UV digestion or other suitable
sample preparation.
Concentration of the analyte is too high: dilute.
Addition solution with metal complexing solution (time reaction).