7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-48
Determination of lead and thallium
With a supporting electrolyte of pH = 1, Pb and Tl peaks overlap greatly.
Changing the pH to 13 separates the two peaks. (The separation of lead
and thallium can also be achieved by subsequent electrolysis or in acetate
buffer with EDTA).
Supporting electrolyte: Supporting electrolyte:
pH = 1 pH = 13
Analysis solution: 0.5 mg/L Pb; 1 mg/L Tl
Standard addition: with 10 µg Pb and 10 µg Tl
Electrode: SMDE
7.5.9 Calibration with chemically non-isoformal standards
With all three possible calibration techniques, it must be ensured that the standards
used for calibration are chemically isoformal with the analytes. The standard sub-
stances must therefore have the same valency (e.g. with Fe, Al) or complex forma-
tion form (e.g. with As, Cr, Se) as the substances already present in the analysis
solution. If this is not the case the calibration can produce completely wrong results
owing to the different peak voltages and sensitivities.
Pb
Tl