Owner manual

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-50
7.6.2 Prepare instrument for diagnostic test
1
Power off.
2
Disconnect all cables at rear panel, except mains cable and connecting cable
to VA Stand.
3
Power on.
7.6.3 Checking thermal printer
1
Prepare instrument for diagnostic test (see section 7.6.2).
2
<VA TRACE ANALYZER>
3
<Diagn.>
4
<Thermoprinter>
5
<Test print>
A test printout is outputted (see following page).
Check printout for completeness and neatness.
The character set (the first five lines) must match the specimen.
The very top and very bottom points of a line must be printed clearly;
check K and p especially.
The printout must have a total length of 260 mm (tolerance: ± 10 mm)
6
Press <ESC> several times until the ”MONITORING” page appears.