7.6.2 Prepare instrument for diagnostic test
1
Power off.
2
Disconnect all cables at rear panel, except mains cable and connecting cable
to VA Stand.
3
Power on.
7.6.3 Checking thermal printer
1
Prepare instrument for diagnostic test (see section 7.6.2).
2
<VA TRACE ANALYZER>
3
<Diagn.>
4
<Thermoprinter>
5
<Test print>
A test printout is outputted (see following page).
• Check printout for completeness and neatness.
• The character set (the first five lines) must match the specimen.
• The very top and very bottom points of a line must be printed clearly;
check K and p especially.
• The printout must have a total length of 260 mm (tolerance: ± 10 mm)
6
Press <ESC> several times until the ”MONITORING” page appears.