3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-34
3.9.2 Tubing connection for addition of standard solutions
For the addition of standard solutions in the µL-range into the measuring vessel of
the 747 VA Stand the 4-way microtip 4747 (6.1824.000) should be used. It is fitted
with 4 lengths of PTFE tubing with connection nipples for direct attachment to the
Exchange unit of the 685 Dosimat. To ready the 685 Dosimat for automatic dis-
pensing, proceed as follows:
1
Mount Exchange unit on 685 Dosimat
Procedure, see section 2 of 685 Instructions for Use.
2
Insert 4-way microtip in 747 VA Stand (see section 3.3.2)
Remove stopper from nipple 4444 and insert 4-way microtip 4747 into nipple
4444 as far as it will go (see Figs 7 and 9).
Tighten nipple 4444 using a 6.2739.000 Wrench (accessory of 6.3014.XXX
Exchange unit) so that the 4-way microtip 4747 can no longer move.
Pull the 4 lengths of PTFE tubing of the 4-way microtip 4747 in succession
from above through the opening 8686.
3
Connect PTFE tubing to Exchange unit
Unscrew the attached 6.1805.100 FEP Tubing
(L = 40 cm) from connection 2 (connection for burette
tip) of the flat stopcock on the Exchange unit mounted
on the Dosimat.
Screw connection nipple of the PTFE tubing of the
4-way microtip 4747 onto connection 2 of the flat stop-
cock on the Exchange unit mounted on the Dosimat.
2
4
Close unused PTFE tubings
Screw a 6.1808.000 Coupling (accessory of 747 VA Stand) on each
unused PTFE tubing of the 4-way microtip 4747.
Screw a 6.1446.040 Dummy stopper (accessory of 747 VA Stand) on
each 6.1808.000 Coupling.
5
Initialize Dosimat(s)
Switch on 746 VA Trace Analyzer using the mains switch 1111.
Select dialog page "DOSIMATS".
If volumes 50 µL should be dispensed, set parameter V.init to
0.05 mL.
Set parameter V.tube in to 1.7 mL.
Set parameter V.tube out to 0.2 mL.
Press <Prime all> for initializing all attached Dosimats or select desired
Dosimat with the cursor keys <ç> or <è> and press <Prime Dosi-
mat>: The burette cylinder is filled and excess solution is expelled to the
measuring vessel. With standard accessories, all tubings to the measur-
ing vessel will have been filled at the end of the initializing procedure.
Start the initializing procedure a second time by pressing the softkey
<Prime all> or <Prime Dosimat> again.
Start the initializing procedure a third time by pressing the softkey <Prime
all> or <Prime Dosimat> again.