Owner manual
3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-34
3.9.2 Tubing connection for addition of standard solutions
For the addition of standard solutions in the µL-range into the measuring vessel of
the 747 VA Stand the 4-way microtip 4747 (6.1824.000) should be used. It is fitted
with 4 lengths of PTFE tubing with connection nipples for direct attachment to the
Exchange unit of the 685 Dosimat. To ready the 685 Dosimat for automatic dis-
pensing, proceed as follows:
1
Mount Exchange unit on 685 Dosimat
• Procedure, see section 2 of 685 Instructions for Use.
2
Insert 4-way microtip in 747 VA Stand (see section 3.3.2)
• Remove stopper from nipple 4444 and insert 4-way microtip 4747 into nipple
4444 as far as it will go (see Figs 7 and 9).
• Tighten nipple 4444 using a 6.2739.000 Wrench (accessory of 6.3014.XXX
Exchange unit) so that the 4-way microtip 4747 can no longer move.
• Pull the 4 lengths of PTFE tubing of the 4-way microtip 4747 in succession
from above through the opening 8686.
3
Connect PTFE tubing to Exchange unit
• Unscrew the attached 6.1805.100 FEP Tubing
(L = 40 cm) from connection 2 (connection for burette
tip) of the flat stopcock on the Exchange unit mounted
on the Dosimat.
• Screw connection nipple of the PTFE tubing of the
4-way microtip 4747 onto connection 2 of the flat stop-
cock on the Exchange unit mounted on the Dosimat.
2
4
Close unused PTFE tubings
• Screw a 6.1808.000 Coupling (accessory of 747 VA Stand) on each
unused PTFE tubing of the 4-way microtip 4747.
• Screw a 6.1446.040 Dummy stopper (accessory of 747 VA Stand) on
each 6.1808.000 Coupling.
5
Initialize Dosimat(s)
• Switch on 746 VA Trace Analyzer using the mains switch 1111.
• Select dialog page "DOSIMATS".
• If volumes ≤ 50 µL should be dispensed, set parameter V.init to
0.05 mL.
• Set parameter V.tube in to 1.7 mL.
• Set parameter V.tube out to 0.2 mL.
• Press <Prime all> for initializing all attached Dosimats or select desired
Dosimat with the cursor keys <ç> or <è> and press <Prime Dosi-
mat>: The burette cylinder is filled and excess solution is expelled to the
measuring vessel. With standard accessories, all tubings to the measur-
ing vessel will have been filled at the end of the initializing procedure.
• Start the initializing procedure a second time by pressing the softkey
<Prime all> or <Prime Dosimat> again.
• Start the initializing procedure a third time by pressing the softkey <Prime
all> or <Prime Dosimat> again.