746 VA Trace Analyzer / 747 VA Stand
3-37
3.9.4 Changing the Exchange unit
The Exchange unit mounted on the 685 Dosimat can be changed only in the
exchange position. Please proceed as follows:
1
Switch to exchange position
Select dialog page "DOSIMATS".
Select desired Dosimat with the cursor keys and press softkey <Exchange
position>: The Dosimat is switched to the exchange position and the
parameter Position is set to exchange.
2
Change Exchange unit
Unscrew connection nipple of the PTFE tubing of the 4-way microtip 4747
from connection 2 and take off old Exchange unit.
Mount new Exchange unit on Dosimat and screw connection nipple of the
PTFE tubing of the 4-way microtip 4747 onto connection 2 of the flat stop-
cock on the Exchange unit.
3
Initialize Dosimat(s)
Switch on 746 VA Trace Analyzer using the mains switch 1111.
Select dialog page "DOSIMATS".
Select desired Dosimat with the cursor keys and press softkey <Prime
Dosimat> for initializing the Dosimat. The burette cylinder is filled and ex-
cess solution is expelled to the measuring vessel. With standard acces-
sories, all tubings to the measuring vessel will be filled at the end of the
initializing procedure.
Start the initializing procedure a second time by pressing the softkey
<Prime Dosimat> again.
Start the initializing procedure a third time by pressing the softkey or
<Prime Dosimat> again.