3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-42
3.10.2 Tubing connection for addition of standard solutions
For the addition of standard solutions in the µL-range into the measuring vessel of
the 747 VA Stand the 4-way microtip 4747 (6.1824.000) should be used. It is fitted
with 4 lengths of PTFE tubing with connection nipples for direct attachment to the
Dosing unit of the 700 Dosino. To ready the 700 Dosino for automatic dispensing,
proceed as follows:
1
Mount Dosing unit on 700 Dosino
Procedure, see section 2.2 of 700 Instructions for Use.
2
Insert 4-way microtip in 747 VA Stand (see section 3.3.2)
Remove stopper from nipple 4444 and insert 4-way microtip 4747 into nipple
4444 as far as it will go (see Figs 7 and 9).
Tighten nipple 4444 using a 6.2739.000 Wrench (accessory of 6.3030.XXX
Dosing unit) so that the 4-way microtip 4747 can no longer move.
Pull the 4 lengths of PTFE tubing of the 4-way microtip 4747 in succession
from above through the opening 8686.
3
Connect PTFE tubing to Dosing unit
Screw connection nipple of the PTFE tubing of the 4-way microtip 4747 onto
connection 1 of the Dosing Unit mounted on the Dosino.
To lengthen the tubing connection the 6.1805.100 FEP tubing (L = 40
cm, accessory of Dosing Unit) can be inserted between PTFE tubing and
connection 1 of the Dosing Unit using the 1808.000 coupling.
4
Close unused PTFE tubings
Screw a 6.1808.000 Coupling (accessory of 747 VA Stand) on each
unused PTFE tubing of the 4-way microtip 4747.
Screw a 6.1446.040 Dummy stopper (accessory of 747 VA Stand) on
each 6.1808.000 Coupling.
5
Initialize Dosino(s)
Switch on 746 VA Trace Analyzer using the mains switch 1111.
Select dialog page "DOSIMATS".
If volumes 50 µL should be dispensed, set parameter V.init to
0.05 mL.
Set parameter V.tube in to 0.7 mL.
Set parameter V.tube out to 0.2 mL.
Press softkey <Prime all> for initializing all attached Dosinos or select
desired Dosino with the cursor keys and press softkey <Prime Dosimat>:
The cylinder is filled and excess solution is expelled to the measuring
vessel. With standard accessories, all tubings to the measuring vessel will
have been filled at the end of the initializing procedure.