Owner manual

3.2 Remote control commands
748 DH Sample Changer
12
3.2 Remote control commands
3.2.1 Overview
Object Description Input range Reference
& Root
à Stirer Stirrer control $G 3.2.2.1.
³ Ã .On Switching stirrer True, False ditto
³
à GoTo Go to position $G 3.2.2.2.
³ Ã .Vessel Designation of vessel special ditto
³
à GoAbove Go to position $G 3.2.2.3.
³ Ã .Vessel Designation of vessel special ditto
³
à GoIn Go into vessel $G 3.2.2.4.
³ Ã .Vessel Designation of vessel special ditto
³
à GoToHome Goes to home position $G 3.2.2.5.
³
à WashElectrodes Wash electrodes $G 3.2.2.6.
³ Ã .Times Number of wash cycles 1...9999 ditto
³
à SetCtrlLine Setting I/O line $G 3.2.2.7.
³ Ã .On Setting line True, False ditto
³ Ã .Line Selection of line 1...3 ditto
³
à WaitForInput Waiting for input on I/O line $G 3.2.2.8.
³ Ã .On Line state True, False ditto
³ Ã .Line Selection of line 1...3 ditto
³
à Init Setup of Sample Changer $G 3.2.2.9.
³ Ã .SpeedXY Speed in XY direction 0...5...7 ditto
³ Ã .SpeedXYFast Speed in XY direction 0...2...7 ditto
³ Ã .SpeedZ Speed in Z direction 0...3...7 ditto
³ Ã .SpeedZFast Speed in Z direction 0...7 ditto
³ Ã .SpeedZSlow Speed in Z direction 0...7 ditto
³
à StopAll Emergency stop $G 3.2.2.10.
³
à Assembly Basic assembly functions
³ Ã .GoToXYZ Go to postion $G 3.2.2.11.
³ ³ Ã .X X coordinate 0...630 ditto
³ ³ Ã .Y Y coordinate 0...272 ditto
³ ³ Ã .Z Z coordinate 0...120 ditto
³ ³ Ã .SpeedXY Speed in XY direction 0...3...7 ditto
³ ³ Ã .SpeedZ Speed in Z direction 0...3...7 ditto
³ Ã .PumpOn Switching on/off pump True, False, $G 3.2.2.12.