User Manual

4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
91
Setup.Lock.Config ON, OFF
Setup.Lock.Parameter ON, OFF
Setup.Lock.SmplData ON, OFF
Setup.Lock.UserMeth.Recall ON, OFF
Setup.Lock.UserMeth.Store ON, OFF
Setup.Lock.UserMeth.Delete ON, OFF
Setup.Lock.Exchange ON, OFF
Setup.Lock.Display ON, OFF
ON means disable the corresponding function:
.Keyboard: Disable all keys of the Coulometer
.Config: Disable the <CONFIG> key
.Parameter: Disable the <PARAM> key
.SmplData: Disable the <SMPL DATA> key
.UserMeth.Recall: Disable "recall" in <USER METH> key
.UserMeth.Store: Disable "store" in < USER METH > key
.UserMeth.Delete: Disable "delete" in < USER METH > key
.Exchange: Disable the <EXCH> key
.Display: Disable the display, i.e. it will not be written to by the device program
of the Coulometer and can be operated from the computer.
4.2.2.90. Setup.Mode.StartWait ON, OFF
Setup.Mode.FinWait ON, OFF
Holding points in the method sequence. If they are "ON", the sequence stops until
"OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method (holding point after AutoInfo
!".T.GC").
.FinWait: Holding point at the end a method (holding point after AutoInfo !".T.F").
4.2.2.91. Setup.SendMeas.SendStatus ON, OFF
Setup.SendMeas.Interval 0.4...4...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
4.2.2.94 and 4.2.2.95) in the inputted interval is active.
.Interval: Time interval (in s) for the automatic transmission of associated meas-
ured values defined under points 4.2.2.95 and 4.2.2.96. The inputted
value is rounded off to a multiple of 0.4. The smallest possible time in-
terval depends on the number of measured values which have to be
sent, on the baud rate, on the load on the interface and on the type of
device connection. With "MPList" the measured values are sent at the
time of their entry into the measured point list.
The automatic transmission is switched on/off with 'SendStatus'.
4.2.2.92. Setup.SendMeas.Select Assembly, Titrator
Selection of the unit of which the measured values should be sent (4.2.2.95 or
4.2.2.96).
4.2.2.93. Setup.SendMeas.Assembly.CyclNo ON, OFF
Setup.SendMeas.Assembly.I ON, OFF
Setup.SendMeas.Assembly.Meas ON, OFF