User Manual

4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
78
Assignment of the statistics calculations. Valid assignments are a requirement
for statistics calculations. In addition, the statistics calculation must be
switched on, see 4.2.2.13. Rules for statistics calculations see page 37.
4.2.2.27. Mode.CFmla
Mode.CFmla.1.Value 0...±999 999
Mode.CFmla.2.Value
etc., up to .19
Calculation constants specific to a method. Stored in the method memory of
the Coulometer. Operands specific to the sample (4.2.2.57 and 4.2.2.59) and
values of common variables (4.2.2.55) on the other hand are not stored with
the methods.
4.2.2.28. UserMeth.FreeMem read only
Memory space, available for user methods or silo lines. $Q sends the number
of free bytes, e.g.
"4928".
4.2.2.29. UserMeth.Recall $G
UserMeth.Recall.Name up to 8 ASCII characters
UserMeth.Store $G
UserMeth.Store.Name up to 8 ASCII characters
UserMeth.Delete $G
UserMeth.Delete.Name up to 8 ASCII characters
UserMeth.DelAll $G
Management of the internal method memory: Load, store and delete methods.
An action is performed if "$G" is sent to the corresponding node just after en-
tering the name.
Do not use blank characters before and after method name!
.DelAll: Deletes all methods in the user memory.
4.2.2.30. UserMeth.List.1.Name read only
UserMeth.List.1.Mode read only
UserMeth.List.1.Bytes read only
UserMeth.List.1.Checksum read only
for each method
List of the methods in the user method memory with the following character-
istics:
.Name: Name of the method
.Mode: Mode
.Bytes: Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 4.2.2.65.
4.2.2.31. Config.Monitoring.Reagent.Status ON, OFF
Config.Monitoring.Reagent.Determ 1...99...999, OFF
Config.Monitoring.Reagent.DCounter 0...999
Config.Monitoring.Reagent.MaxTime 1...7...9999, OFF
Config.Monitoring.Reagent.TCounter 0...9999