4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
85
.EP: Endpoint:
Mass coordinate in ug, e.g. "10.3"
Potential coordinate in mV e.g. "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
4.2.2.66.
C40: Initial measured value in mV, e.g. "226".
C41: Mass of water in ug, e.g. "126.5"
C42: Time from start of titration to end in s, e.g. "26".
C43: Drift at titration start in ug/min, e.g. "5.1".
C44: Titration temperature in °C, e.g. "25.0"
C45: Total charge in mA·s, e.g. "1355.5"
4.2.2.67. Info.StatisticsVal.ActN read only
Info.Statistics.1.Mean read only
Info.Statistics.1.Std read only
Info.Statistics.1.RelStd read only
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results "3"
Data for MN1:
Mean: Mean value (decimal places as in result) "3.421"
Std: Standard deviation (1 decimal place more than in result) "0.0231"
RelStd: Relative standard deviation (in %, 2 decimal places) "0.14"
4.2.2.68. Info.SiloCalc.C24.Name read only
Info.SiloCalc.C24.Value read only
Info.SiloCalc.C24.Unit read only
for .C25 as for .C24
Info.SiloCalc.C26.ActN read only
Info.SiloCalc.C26.Mean read only
Info.SiloCalc.C26.Std read only
Info.SiloCalc.C26.RelStd read only
for .C27 as for .C26
The current values from the silo calculations. C26 is the mean value out of the
C24 variables; C27 comes from C25.
$Q sends:
C24.Name: Name of the assigned value "RS1"
C24.Value: Value "2.222"
C24.Unit: Unit of the assigned value "%"
C26.ActN: Number of single results "3"
C26.Mean: Mean (decimal places as for the result itself) "3.421"
C26.Std: Standard deviation (decimal places as for the result + 1) "0.0231"
C26.RelStd: Relative standard deviation (in %, 2 decimal places) "0.14"
4.2.2.69. Info.ActualInfo.Inputs.Status read only
Info.ActualInfo.Inputs.Change read only
Info.ActualInfo.Inputs.Clear $G
Info.ActualInfo.Outputs.Status read only
Info.ActualInfo.Outputs.Change read only