4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
90
4.2.2.83. Assembly.Bur.Fill $G, $H, $C
$G starts the 'FILL' mode of the connected Dosino.
4.2.2.84. Assembly.Bur.ModeDis $G, $S, $H, $C
Assembly.Bur.ModeDis.Selected volume, time
Assembly.Bur.ModeDis.V 0.0001...0.1...9999
Assembly.Bur.ModeDis.Time 0.25...1...86400
Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF
Assembly.Bur.ModeDis.AutoFill ON, OFF
Dispensing mode for the connected Dosino. The dispensing mode can only be
started and stopped via the RS Control. During a running dosification, no
method can be started at the Coulometer.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop: Limit volume for the dispensing.
.AutoFill: ON means automatic filling after every dispensing.
4.2.2.85. Setup.Comport 1, 2, 1&2
Selects the Coulometer COM for the output of automatic info:
&Setup.Keycode
&Setup.Trace
&Setup.SendMeas
&Setup.AutoInfo
4.2.2.86. Setup.Keycode ON, OFF
ON means the key code of a key pressed on the Coulometer is outputted. The
key code comprises 2 ASCII characters; table of the keys with their code, see
page 96. A keystroke of key 11 is sent as follows:
#11
The beginning of the message is marked by a space (ASCII 32).
4.2.2.87. Setup.Tree.Short ON, OFF
Setup.Tree.ChangedOnly ON, OFF
Definition of the type of answer to $Q.
.Short: With "ON", each path is sent with only the necessary amount of
characters in order to be unequivocal (printed in bold in this man-
ual). A combination of .Short and .ChangedOnly is not possible.
.ChangedOnly: Sends only the changed values, i.e. values which have been
edited. All paths are sent absolute, i.e. from the root.
4.2.2.88. Setup.Trace ON, OFF
The Coulometer automatically reports when a value has been confirmed with
<ENTER> at the Coulometer. Message, e.g.:
&SmplData.OFFSilo.Id1"Trace"
The beginning of the message is marked by a space (ASCII 32).