Owner manual

3.8 Connection of 765 Dosimats
757 VA Computrace – Hardware
41
3.8.2 Tubing connection
For the addition of standard or auxiliary solutions into the measuring vessel of the
757 VA Computrace Stand the 4-way microtip
26
(6.1824.000) can be used. It is fit-
ted with 4 lengths of PTFE tubing with connection nipples for direct attachment to
the Exchange unit of the 765 Dosimat. To ready the 765 Dosimat for automatic dis-
pensing, proceed as follows:
1 Mount Exchange unit on 765 Dosimat
Procedure, see section
5 of 765 Instructions for Use
.
2 Insert 4-way microtip in 757 VA Computrace Stand
(see
section
3.2.4
)
Remove stopper from nipple
23
and insert 4-way microtip
26
into nipple
23
as far as it will go (see
Figs 3 and 6
).
Tighten nipple
23
using a 6.2739.000 Wrench (accessory of 6.3014.XXX
Exchange unit) so that the 4-way microtip
26
can no longer move.
Pull the 4 lengths of PTFE tubing of the 4-way microtip
26
in succession
from above through the opening
65
.
3 Connect PTFE tubing to Exchange unit
Unscrew the attached 6.1805.100 FEP Tubing
(
L
= 40 cm) from connection
2
(connection for burette
tip) of the flat stopcock on the Exchange unit mounted
on the Dosimat.
Screw connection nipple of the PTFE tubing of the
4-way microtip
26
onto connection
2
of the flat stop-
cock on the Exchange unit mounted on the Dosimat.
2
4 Close unused PTFE tubings
Screw a 6.1808.000 Coupling (accessory of 757 VA Computrace Stand)
on each unused PTFE tubing of the 4-way microtip
26
.
Screw a 6.1446.040 Dummy stopper (accessory of 757 VA Computrace
Stand) on each 6.1808.000 Coupling.
5 Initialize Dosimat(s)
Switch on the 765 Dosimat.
Switch on the 757 VA Computrace Stand using the mains switch
12
.
Switch on the PC and start the VA Computrace program.
Click on
or
MAIN WINDOW / Utility / Dosimat control
to open the
DOSI-
MAT CONTROL
window.
Select the desired Dosimat in the
Burette
field.
Click the
button to empty and refill the exchange unit in-
stalled on the Dosimat. Repeat this procedure two times.
Check if there are air bubbles left in the glass cylinder of the exchange
unit. If this is the case, repeat the flushing procedure by clicking the
button.
Close the
DOSIMAT CONTROL
window.