Owner manual

3 Installation
757 VA Computrace – Hardware
48
3.9.3 Software settings
Before putting into operation the 757 VA Computrace Stand with the 813 Compact
Autosampler, the following settings have to be made in the "
757 VA Computrace
2.0
" software program:
1 Set Dosimat parameters
Click on
MAIN WINDOW
/ Settings / General settings
and select the
Hardware
tab.
For each Dosimat connected to the remote interface of the 757 VA Com-
putrace Stand, check the
Dosimat no.
checkbox and enter the
Volume
of
the exchange unit installed on the Dosimat and the
Dose rate
of the Dosi-
mat.
Define the addition or predose solutions for the desired method in the
DOSIMATS
window (procedure see
Software Manual,
section 5.2
).
Fill the exchange units of the Dosimats with the desired solutions and
make sure that there are no gas bubbles in the glass cylinders using the
button in the
DOSIMAT CONTROL
window (see
section 6.2
,
Software Manual
).