Owner's manual

2.7 Settings menu
757 VA Computrace – Software
23
Dosimat no. [ on, off; off ]
Check this checkbox for each Dosimat con-
nected.
Volume (mL) [ 1, 5, 10, 20, 50 mL; 5 mL ]
Volume of the exchange unit of the Dosimat
Dose rate (mL/min) [ 0.001 ... 150 mL/min (depen-
ding on exchange unit) ]
Dosing rate of the Dosimat.
Note: If Dosimats are connected to the 757 VA Computrace
Stand, the instruments must always be switched on in the se-
quence Dosimat 757 PC.
Automation
757 VA COMPUTRACE / Settings / General Settings
In the
GENERAL SETTINGS
window default set-
tings for the operation of the 813 Compact
Autosamplers and two 772 Pump Units can
be defined with the
Automation
tab.
Use Autosampler
Enable/disable the use of the 813 Compact
Autosampler. If this option is disabled, the
SAMPLE TABLE
window will not be available.