2 Main window
757 VA Computrace – Software
24
Note:
Method 2
must be set on the 813 Compact
Autosampler.
Time to move to sample (s) [ 30 s ; 25 ... 2147483647 s ]
Maximum time needed for the 813 Compact
Autosampler to move from one sample to the
other. (minimum time allowed:
25 s
).
Sample transfer time (s) [ 300 s ; 30 ... 2147483647 s ]
Time to transfer the sample solution from the
sample vessel to the measurement vessel using
the peristaltic pump of the 813 Compact Auto-
sampler (minimum time allowed:
30 s
).
No. of rinsing cycles [ 3 ; 0 ... 2147483647 ]
Number of rinsing cycles.
Siphoning time (s) [ 25 s ; 0 ... 2147483647 s ]
For each rinsing cycle, the measuring vessel is
siphoned off during this time using a 772 Pump
Unit.
Rinsing time (s) [ 5 s ; 0 ... 2147483647 s ]
For each rinsing cycle, the measuring vessel is
rinsed during this time using a 772 Pump Unit.
Test the 813 Compact Autosampler with the set
automation parameters.
Note: Before starting the test, switch on the 813 Compact Auto-
sampler, set
Method 2
at the Autosampler and place two sample
vessels filled with water on the sample rack.
For details on the use of the 813 Compact Autosampler, see 813
Instructions for Use.
Save settings
MAIN WINDOW / Settings / Save now
This function saves the actual settings of the
software: Open windows, window position and
size, general settings.
MAIN WINDOW / Settings / Save on exit
If this function is enabled, the software settings
are stored when the software is quit.