757 VA Computrace – Software
28
Voltage step time
t
Hg drop
surface
Notes:
In the exploratory mode, the DME can be used for all meas-
urement modes except SqW, CV and PSA. In the determina-
tion mode, the DME can be used for all measurement modes
except DC, SqW, CV and PSA.
An advantage of the DME compared with the SMDE is that
the MME capillary is subjected to less mechanical stress.
A disadvantage of the DME compared with the SMDE and
HMDE is the higher mercury consumption and the lower sen-
sitivity as the electrode surface constantly changes during the
measurement phase.
SMDE
SMDE is an electrode mode of the Multi-Mode Electrode and
stands for Static Mercury Drop Electrode. It combines the fea-
tures of the DME and the HMDE: as with the DME, the mercury
drops are constantly renewed, but during the measurement the
drop surface is constant as in the HMDE case. Each mercury
drop is knocked off by a tapping mechanism after the
Voltage step
time
set in the measurement mode.
Drop size
[ 1...9 ; 4 ]
Size of the mercury drop (surface 0.15 mm
2
...0.60 mm
2
).
t
Voltage step time
Drop size
×
40 ms
Hg drop
surface