3.2 Remote control commands
758 KFD Titrino
113
3.2.2.79. Config.RSSet1 $G
Config.RSSet1.Baud 300, 600, 1200, 2400, 4800, 9600,
19200, 38400, 57600, 115200
Config.RSSet1.DataBit 7, 8
Config.RSSet1.StopBit 1, 2
Config.RSSet1.Parity even, odd, none
Config.RSSet1.Handsh HWs, SWchar, SWline, none
$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 130ff.
Baud rates >9600 need a PC which is equipped accordingly (e.g. with 16550
component).
The setting of the values must be initiated with $G immediately after entry of
the values.
3.2.2.80. Config.ComVar.C30
with up to .C39, etc. 0... ±999999
Values of the common variables from C30 up to C39. Insert the common
variables directly or describe the determination results directly from the
method, see 3.2.2.7.
3.2.2.81. Config.DosPrep.PowerOnPrep ON, OFF
Warning after power on, that the preparation of the buret has to be executed.
3.2.2.82. Config.DosPrep.Report ON, OFF
Report output after preparation of burets.
3.2.2.83. Config.DosPrep.Select internal D0, external D1, external D2
Selection of buret.
3.2.2.84. Config.DosPrep.DX.WarnInterval 5...9999, OFF
Time interval in min for automatic warning that a preparation has to be exe-
cuted.
3.2.2.85. Config.DosPrep.DX(.Dosimat).V 0...3.5...99999.99
Volume in mL, which will be expelled at the preparation. Start sequence see
3.2.2.121.
3.2.2.86. Config.DosPrep.DX(.Dosimat).Repeat 1...2...9
Number of cycles for expelling the volume at the preparation. Start sequence
see 3.2.2.121.