3.2 Remote control commands
758 KFD Titrino
114
3.2.2.87. Config.DosPrep.DX(.Dosimat).DosRate 0.01...150, max.
Config.DosPrep.DX(.Dosimat).FillRate 0.01...150, max.
Dosing and filling rate in mL/in for the preparation. Start sequence see
3.2.2.121.
3.2.2.88. Config.DosPrep.DX.Select Dosimat, Dosino
Selection of buret type. Start sequence see 3.2.2.121.
3.2.2.89. Config.DosPrep.DX.Dosino.Outlet tip, flask
Place where the liquid of the preparation is expelled. Start sequence see
3.2.2.121.
tip: To the tip.
flask: Back to the flask.
3.2.2.90. Config.DosPrep.DX.Dosino.DosTubing.Length 1...40...999.9
Config.DosPrep.DX.Dosino.DosTubing.Diameter 0.1...2...9.9
Config.DosPrep.DX.Dosino.AspirTubing.Length 1...25...999.9
Config.DosPrep.DX.Dosino.AspirTubing.Diameter 0.1...2...9.9
.DosTubing: Dosing tube. Length in cm, diameter in mm.
.AspirTubing: Aspiration tube. Length in cm, diameter in mm.
Start sequence see 3.2.2.121.
3.2.2.91. Config.DosPrep.DX.Dosino.DosRate 0.01...150, max.
Config.DosPrep.DX.Dosino.FillRate 0.01...150, max.
Dosing and filling rate in mL/min for preparation. Start sequence see
3.2.2.121.
3.2.2.92. SmplData.Status ON, OFF
On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.
3.2.2.93. SmplData.OFFSilo.Id1 up to 8 ASCII characters
SmplData.OFFSilo.Id2 up to 8 ASCII characters
SmplData.OFFSilo.Id3 up to 8 ASCII characters
SmplData.OFFSilo.ValSmpl 6-digits, sign and decimal point
SmplData.OFFSilo.UnitSmpl up to 5 ASCII characters
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific calcu-
lation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.